Mechanically ventilated patients require appropriate settings of respiratory control variables to maintain acceptable gas exchange. To control the carbon dioxide (CO2) level effectively and automatically, system identification based on a human subject was performed using a linear affine model and a nonlinear Hammerstein structure. Subsequently, a robust controller was designed using the H∞ loop-shaping approach, which synthesizes the optimal controller based on a specific objective by achieving stability with guaranteed performance. For demonstration purposes, the closed-loop control ventilation system was successfully tested in a human volunteer. The experimental results indicate that the blood CO2 level may indeed be controlled noninvasively by measuring end-tidal CO2 from expired air. Keeping the limited amount of experimental data in mind, we conclude that H∞ loop-shaping may be a promising technique for control of mechanical ventilation in patients with respiratory insufficiency.
ROBUST CONTROL OF END-TIDAL CO2 USING THE H∞ LOOP-SHAPING APPROACH
2013-12-31
doi:10.14311/AP.2013.53.0895
Acta Polytechnica; Vol 53, No 6 (2013); 895-900 ; 1805-2363 ; 1210-2709