Automated apparatus reveals sources of error not revealed in calibration. Computer-controlled apparatus detects and measures Seebeck inhomogeneities in sheathed thermocouples. Measures thermocouple output voltage as function of position of probe along sharp gradient of temperature. Abnormal variations in voltage-versus-position data indicative of Seebeck inhomogeneities. Prototype for development of standard method and equipment for routine acceptance/rejection testing of sheathed thermocouples in industrial and research laboratories.


    Access

    Access via TIB

    Check availability in my library


    Export, share and cite



    Title :

    Apparatus Tests Thermocouples For Seebeck Inhomogeneity



    Published in:

    Publication date :

    1995-01-01



    Type of media :

    Miscellaneous


    Type of material :

    No indication


    Language :

    English