A description of the precision triaxial fluxgate magnetometer to be flown aboard the MAGSAT spacecraft is presented. The instrument covers the range of + or - 64,000 nT with a resolution of + or - 0.5 nT, an intrinsic accuracy of + or - 0.001% of full scale and an angular alignment stability of the order of 2 seconds of arc. It was developed at NASA's Goddard Space Flight Center and represents the state-of-the-art in precision vector mangetometers developed for spaceflight use.


    Zugriff

    Zugriff über TIB

    Verfügbarkeit in meiner Bibliothek prüfen


    Exportieren, teilen und zitieren



    Titel :

    The MAGSAT Vector Magnetometer: A Precision Fluxgate Magnetometer for the Measurement of the Geomagnetic Field


    Beteiligte:
    M. H. Acuna (Autor:in) / C. S. Scearce (Autor:in) / J. Seek (Autor:in) / J. Scheifele (Autor:in)

    Erscheinungsdatum :

    1978


    Format / Umfang :

    48 pages


    Medientyp :

    Report


    Format :

    Keine Angabe


    Sprache :

    Englisch





    The Magsat magnetometer boom