An interlaboratory comparison of Josephson Voltage Standards (JVS) has been made among 16 national, industrial; and military standards laboratories in North America and one in Europe. The comparison was made at 10 V using a set of four traveling Zener reference standards. A pivot laboratory made measurements at the beginning, at the end, and at nine other times during the comparison. The measured differences and their uncertainties are reported and used to establish a table of equivalence between each participant and the pivot, and between each participant and the National Institute of Standards and Technology (NIST). All but two of the differences fall within two parts in 108.


    Zugriff

    Zugriff über TIB

    Verfügbarkeit in meiner Bibliothek prüfen

    Bestellung bei Subito €


    Exportieren, teilen und zitieren



    Titel :

    Interlaboratory comparison of Josephson voltage standards


    Beteiligte:
    Deaver, D. (Autor:in) / Miller, W.B. (Autor:in) / Pardo, L. (Autor:in) / Jaeger, K. (Autor:in) / Plowman, D. (Autor:in) / Hamilton, C.A. (Autor:in)


    Erscheinungsdatum :

    2001


    Format / Umfang :

    4 Seiten, 7 Quellen




    Medientyp :

    Aufsatz (Konferenz)


    Format :

    Print


    Sprache :

    Englisch






    Straylight of curved gratings: an interlaboratory comparison

    Kroneberger, Monika / Tomuta, Dana / Mezger, Andreas | British Library Conference Proceedings | 2019



    Proposed Josephson voltage standard