Electrical measurements made by four point probes, two of which double as temperature probes. Laboratory apparatus measures both Seebeck coefficients and electrical resistivities of candidate thermoelectric materials at temperatures from ambient to 1,300 K. Apparatus makes possible to take both measurements alternately and in rapid succession during same heating cycle, thereby reducing distortion.
Apparatus Measures Seebeck Coefficient And Resistivity
NASA Tech Briefs ; 17 , 5
01.05.1993
Sonstige
Keine Angabe
Englisch
Determination of the Seebeck Coefficient of Thin Film Thermocouples
British Library Conference Proceedings | 1998
|Microstructures for characterization of seebeck coefficient of doped polysilicon films
British Library Online Contents | 2011
|Apparatus Tests Thermocouples for Seebeck Inhomogeneity
Online Contents | 1995
Apparatus Tests Thermocouples For Seebeck Inhomogeneity
NTRS | 1995
|